Technical Data

Measurement using direct voltage with a voltage source of 0.61V DC and a source resistance of 61.5 kOhm.

The maximum DC patient auxiliary current according to EN-60601 medical device standard is thus within the limit value of 10uA DC.

Sampling frequency: 5 Hz (5 measured values per second). The measured values are saved for exporting the data and can be exported as a CVS file in the app.

Resolution of the measured values: 18 bit, rounded to 2 decimal places, i.e. 0.01uS in data export without rounding.

Measuring range: 10 kilo-ohms to 1 mega-ohm corresponding to 100uS .. 1uS. Possible deviation of + – 5% over the entire measuring range on average.